Three-circle X-ray diffractometer DSO-2-01

Automated three-circle X-ray diffractometer DSO-2-01 (fig.1) is intended for the automatic measurement of a reflection Bragg angle from a single crystal. Diffractometer allows the measurements of the submillimetric samples with arbitrarily shape, for instance of IMTK crystals (IMTK - Russian abbreviation for the maximum temperature crystal measurement).

Fig.1. General view of DSO-2-01.

Specification

Dimensions

1200õ1000õ1900 (height)  mm

Weight

< 350 kg

Power feeding

220 V 50 – 60 Hz

Electrical power consumption

2500 W

X-ray generator power

1200 W

X-ray tube

0,9BSW29  Fe ( Cu, Cr )

X-ray tube cooling

sity water, thermo-chiller

High voltage

10 – 50 kV

X-ray tube current

5 - 24 mA

Protection cabinet

Steel, lead rubber 3 mm, X-ray protection glass 10 mm

Sample dimensions

diameter 0.2 - 28 mm, thickness 0.1 - 2 mm

2Theta (detector) angle range, deg.

-100 ... +169 (motorized by a stepping motor)

Theta (sample) angle range, deg.

360 (motorized by a stepping motor)

Phi attachment (sample) angle range, deg.

360 (motorized by a stepping motor)

Half step of Theta axis, deg.

0.00025

Half step of 2Theta axis, deg.

0.0005

Half step of Phi axis, deg.

0.1

Maximum rotation speed, rev/sec

1

X (linear detector stage), mm

100 (motorized by a stepping motor)

Half step of X, mm

0.005

Y (linear attachment stage), mm

±4 (motorized by a stepping motor)

Half step of Y, mm

0.0025

Angle encoder increment on Theta and 2Theta, deg.

0.0005

Automatic rotary switch of slits in front of a detector

3 positions (motorized by a stepping motor)

Searching angle range, deg.

90

Scintillation counter

entrance window diameter 24 mm, count rate 6*105 cps

Measurement circle duration

about 15 min.

Software

Radicon Device Programming Workbench (RDPW)

Radiation protection

less than 1 mSv/h (diffractometer  is fully exempted from radiation inspection)

Diffractometer arrangement and operation

        Not monochromatic radiation is used in DSO-2-01, but a beta-filter is on a diffracted beam. X-ray optic diagram is presented in fig.2. Scintillation detector with a circle aperture ±6° should be moved to 2Theta position for a selected reflection, and a sample on the attachment rotates both around Theta-axis and around Phi-axis in a definite Theta angle range. When the reflection is detected, a horizontal slit is set automatically in front of detector, and a Phi position of sample is specified. Theta position is specified by the Omega scanning. Then detector moves automatically to the maximum distance, a narrow vertical slit is set in fronta of it, and a maximum intensity is specified by the detector rotation. The last Theta-2Theta scanning gives a 2Theta angle to be found.

       A general view of three-circle goniometer is shown in fig.3. Goniometer is mounted inside of protection cabinet. The control block and a high voltage generator are mounted inside of diffractometer sceleton. USB or Ethernet channel are used for the diffractometer control.

Fig.2. X-ray optic diagram of DSO-2-01.

Fig.3. Three-circle X-ray goniometer.

 