X-Ray Scintillation Detector SCSD-4

SCSD-4 (Single Channel Scintillation Detector) - an optimum solution for count rate measurements of X-ray photon flux in various types of diffraction, spectral, absorption and other experiments under control of PC computer.

Specification

Energy range, keV

5 - 30

Counting rate, s-1

1 -6*105

Maximum number of counts

4 x 109

Exposure time, s

0.1-500

Noise level at 6 keV (discriminated), cps

< 1

Discriminator threshold range, mA

0 - 2.5

Discriminator threshold resolution, mA

0.0025

Diameter of detector mounting part, mm

40

Diameter of entrance window, mm

24

Connector cable length, m

5 - 25

Weight of detector head, g

700

Features:  

  • Compact, low-consuming, and flexible design, 
  • High-speed counting, wide dynamic range, 
  • Low-voltage detector block connection, 
  • Full computer control of high voltage and discrimination thresholds, 
  • Sound monitor of the X-ray flux with piezo-loudspeaker, 
  • Energy dispersive or integral counting, 
  • Preset time, preset counts or continuous count rate (intensity) measurement, 
  • Quick and easy optimization of HV and threshold values with SPECTRUM utility, 
  • Single measurement or cycled operation, 
  • Optional -- synchronization of external devices with use of BUSY and INHIBIT lines. 

Design:
 
SCSD-4 consists of two blocks:

  • detector head, containing 2 mm thick scintillation NaI(Tl) crystal in assembly with beryllium 24 mm diameter entrance window, photomultiplier, remote-controlled high-voltage converter, preamplifier, pulse-forming circuits, amplitude discriminator, output amplifier;

  • control block, plugging into IBM PC AT COM port (or Ethernet - by the order), containing microprocessor controller and sound monitor.

Detector heads are produced in two versions: for using in open air and in vacuum chambers. Vacuum variant has vacuum-tight housing, inner part of which is connected with atmosphere via two flexible tubes. The detector head has a 15-pin SUB-D cable connector.

Software:

The control software SCSW operates under Windows of current version. It includes: microprocessor loadable program, SCSW server, containing communication and control subroutines, DLL library for data/command interchange between the server and user application program, sample user application program, supplied together with the source codes to help the user in creation of his own applications. Software has nice and friendly user interface, controls up to eight SCSD-4 devices running simultaneously, supports three basic measurement modes, single and cycle measurements, comprises SPECTRUM utility aimed to optimize quickly the detector settings. Modifications of software are available on consumer request.

 

The user interface of the software

 